Data & References

  1. The Bone Marrow MSC Separation Device is ...
  2. How to Use
  3. Data & References
  4. Product Information

Data 3.Surface Marker Analysis

MSCs separated & harvested by the Bone Marrow MSC Separation Device were cultivated and surface markers were analysed at 11 days post-cultivation. The results showed that the cells harvested expressed MSC markers (CD29; CD73; CD90; CD105; CD166), as do the cells harvested by density dependent centrifugation.
Also the cells were negative for monocyte (CD14); haematopoietic cell (CD34); leukocyte (CD45) markers.

Surface markers present on harvested MSCs

[Illustrated] Surface markers present on harvested MSCs